3280-10 CLAMP ON HiTESTER--HIOKI 3280-10 |
|
Product Information
|
Sorry£¬no relevant downloads£¡£¡£¡
|
Easy operation ! -1000 A , clamp aperture: 33 mm(1.30 in) dia. -Light 100 g(3.5 oz), and Slim 16 mm(0.63 in) -Independent-opening double-lever design -Slim body design, allows easy clamping even for narrow conductors -No metal (iron core) exposure, ensuring enhanced safety
Basic specifications |
 |
AC Current range |
42.00 to 1000 A AC, 3 ranges (¡À1.5 % rdg. ¡À dgt. at 50 or 60Hz), Average rectifier effective value |
AC Voltage range |
4.200 to 600 V AC, 4 ranges (¡À2.3 % rdg. ¡À dgt. at 50 or 60Hz), Average rectifier effective value |
DC Voltage range |
420.0 mV to 600 V DC, 5 ranges (¡À1.3 % rdg. ¡À4 dgt.) |
Other functions |
Resistance: 420.0 to 42.00 M-ohm, 6 ranges, Accuracy: ¡À2.0 % rdg. ¡À4 dgt. (at 420 to 420 k-ohm range), Continuity: 420.0 ohm (Buzzer sounds at approx. 50 ohm or less) |
Analog output |
None |
Frequency characteristics (at AC current / voltage) |
AC current: 50/60 Hz, AC voltage: 50 to 500 Hz |
Display section |
Digital / LCD, maximum 4199 dgt. |
Sampling rate |
2.5 times/sec or 1 times/3 sec |
Max. circuit voltage |
600 V AC rms (insulated wire) |
Power supply |
CR2032 (3 VDC) ¡Á 1, Continuous use 150 hours |
Core jaw dia. |
33 mm(1.30 in) |
Dimensions, mass |
57 mm(2.24 in)W ¡Á175 mm(6.89 in)H ¡Á16 mm(0.63 in)D, 100 g (22.44 oz) |
Accessories |
9208 TEST LEAD (1), 9398 CARRYING CASE (1) | |
|
[ top ]
|
|